Tuam Tshoj X-RAY TUBE EQUIVALENT TOSHIBA E7242 Chaw tsim tshuaj paus thiab tsum | Sailray
X-RAY Tube sib npaug rau TOSHIBA E7242

X-RAY Tube sib npaug rau TOSHIBA E7242

X-RAY Tube sib npaug rau TOSHIBA E7242

Lus piav qhia luv luv:

Daim ntawv thov: X-ray raj sib dhos rau txhua qhov kev kuaj mob niaj hnub nrog cov pa
los yog digital radiographic thiab fluoroscopic workstations
◆ Lub ntsiab nta: 12.5 ° Rhenium-Tungsten molybdenum phiaj (RTM)
◆ Focal me ntsis: me 0.6, Loj: 1.2
◆ Qhov siab tshaj qhov hluav taws xob: 125kV
◆ Haum nrog IEC60526 hom high-voltage cable receptacles
◆ Lub tshuab hluav taws xob siab yuav tsum ua raws li IEC60601-2-7
◆IEC Classification (IEC 60601-1: 2005): Class I ME EQUIPMENT

Product Detail

Cov nqe lus them nqi & Shipping:

Khoom cim npe

Standard Reference

Standard Reference

Cov npe

EN 60601-2-54: 2009 Cov khoom siv hluav taws xob kho mob - Ntu 2-54: Cov kev cai tshwj xeeb rau kev nyab xeeb thiab kev ua haujlwm tseem ceeb ntawm cov khoom siv xoo hluav taws xob rau hluav taws xob thiab xov tooj cua
IEC 60526: Cov High-voltage cable plug thiab socket txuas rau cov khoom siv kho mob X-ray
IEC 60522: 1999 (TSI) Kev txiav txim siab mus tas li pom ntawm X-ray raj sib dhos
IEC 60613-2010 1 Hluav taws xob, thermal thiab loading yam ntxwv ntawm rotating anode X-ray raj rau kev kuaj mob
IEC 60601-1: 2006 Cov khoom siv hluav taws xob kho mob - Ntu 1: Cov kev cai dav dav rau kev nyab xeeb thiab kev ua haujlwm tseem ceeb
IEC 60601-1-3: 2008 Cov cuab yeej siv kho mob hluav taws xob - Ntu 1-3: Cov kev cai dav dav rau kev nyab xeeb thiab kev ua haujlwm tseem ceeb - Cov Qauv Txheej Txheem: Kev tiv thaiv hluav taws xob hauv cov khoom siv kuaj X-ray
IEC 60601-2-28: 2010 Cov khoom siv hluav taws xob kho mob - Ntu 2-28: Cov kev cai tshwj xeeb rau kev nyab xeeb thiab kev ua haujlwm tseem ceeb ntawm X-ray raj sib dhos rau kev kuaj mob
IEC 60336-2005 1 Cov khoom siv hluav taws xob kho mob-X-ray raj sib dhos rau kev kuaj mob-Cov yam ntxwv ntawm cov chaw focal

Kev piav qhia

● Lub npe yog tsim raws li hauv qab no:

MWM 7110A

Tube

A

High voltage socket nrog 90 degree kev taw qhia

MWTX 71-0.6/1.2-125

B

High voltage socket nrog 270 degree kev taw qhia

Technical Data

Khoom

Specification

Txuj

Nominal input zog (s) ntawm lub anode

F 1

F 2

TS EN 60613

20kW (50/60Hz)

40kW (50/60Hz)

 

Anode tshav kub cia muaj peev xwm

110 kJ (150 kHU)

TS EN 60613

Qhov siab tshaj plaws txias peev xwm ntawm lub anode

500 W

 
Thaum tshav kub kub cia muaj peev xwm

9 00kj

 
Max. nruam kub dissipation yam tsis muaj cua-cim

180 W

 
Cov khoom siv anodeAnode sab saum toj txheej khoom

Rhenium-Tungsten-TZM (RTM)

Rhenium-Tungsten-(RT)

 
Lub hom phiaj (Ref: reference axis)

12.5° Nws

IEC 60788 (ib.)

X-ray tube los ua ke inherent filtration

1.5 hli Al / 75kV

IEC 60601-1-3: Kev qhia tshwj xeeb

Focal Spot nominal tus nqi

F1 (me me tsom)

F2 (loj tsom xam)

IEC 60336 (TSI)

0.6

1.2

 
X-ray tube nominal voltageRadiographicFluoroscopic

125 kV ua

100kV ua

TS EN 60613

Cov ntaub ntawv ntawm cathode cua sov Max. tam sim no

Max voltage

≈ / AC, <20 kHz

 

F1

F 2

 

5.1A

≈79V

5.1 A

≈1214 V

 
Cov hluav taws xob tawg ntawm 150 kV / 3mA hauv 1m nrug

0.5mGy/h

IEC 60601-1-3

Qhov siab tshaj plaws hluav taws xob

443 × 443mm ntawm SID 1m

 
X-ray raj sib dhos hnyav

Kwv yees li. 18 kg

 

Cov xwm txheej rau kev ua haujlwm, khaws cia thiab thauj

txwv

Kev txwv kev ua haujlwm

Kev thauj mus los thiab kev txwv cia

Ambient kub

Los ntawm 10rau 40

Los ntawm - 20to 70

txheeb ze humidity

≤ 75%

≤ 93%

Barometric siab

Los ntawm 70kPa mus rau 106kPa

Los ntawm 70kPa mus rau 106kPa

 

Stator tseem ceeb qhov tseem ceeb

1-theem stator

Test point

C-M

C-A

Winding kuj

≈18.0…22.0Ω

≈45.0…55.0Ω

Max.permissible khiav hauj lwm voltage (run-up)

230V ± 10% Tawm tswv yim

Pom zoo ua haujlwm voltage (run-up)

160V ± 10% Tawm tswv yim

Braking voltage

70vdc ua

Khiav-on voltage nyob rau hauv raug

80 vwm

Kev khiav hauj lwm voltage hauv fluoroscopy

20V-40Vrms

Lub sijhawm ua haujlwm (nyob ntawm qhov pib qhov system)

1.2s ib

Ceev faj hauv kev ua haujlwm

1 .X-ray Radiationkev tiv thaiv

Cov khoom no ua tiav cov kev cai ntawm IEC 60601-1-3.

X-ray raj sib dhos no tso tawm cov hluav taws xob xoo hluav taws xob hauv kev ua haujlwm.Tsuas yog cov neeg ua haujlwm tsim nyog thiab cov neeg ua haujlwm tau txais kev tso cai ua haujlwm X-ray raj sib dhos.

Cov teebmeem physiologic cuam tshuam yuav ua rau muaj kev phom sij rau tus neeg mob, kev tsim khoom yuav tsum tau txais kev tiv thaiv kom tsis txhob muaj hluav taws xob ionization.

2. Dielectric 0 il

X-ray raj sib dhos muaj dielectric 0il muaj rau kev ruaj ntseg siab. Raws li nws yog tshuaj lom rau tib neeg noj qab haus huv,yog tias nws raug rau thaj tsam tsis txwv,nws yuav tsum muab pov tseg raws li kev cai hauv zos.

3 .Operation Atmosphere

X-ray raj sib dhos tsis raug tso cai siv nyob rau hauv cov huab cua ntawm cov nplaim taws lossis cov roj corrosive ·

4.Kho lub raj tam sim no

Nyob ntawm cov xwm txheej ua haujlwm,cov yam ntxwv filament yuav raug hloov.

Qhov kev hloov pauv no tuaj yeem ua rau muaj kev cuam tshuam ntau dua rau X-ray raj sib dhos.

Txhawm rau tiv thaiv X-ray tube los ua kom puas,kho lub raj tam sim no tsis tu ncua.

Dhau li ntawm thaum X-ray raj muaj teeb meem arcing hauv iblsiv sijhawm ntev,yuav tsum tau kho lub raj tam sim no.

5.X-ray Tube Vaj Tse Kub

Tsis txhob kov ntawm X-ray raj lub tsev nyob saum npoo tsuas yog tom qab ua haujlwm vim qhov kub thiab txias.

Nyob twj ywm X-ray raj kom txias.

6.Txoj kev txwv

Ua ntej siv,thov paub meej tias qhov xwm txheej ib puag ncig yog nyob rau hauv kev khiav hauj lwm Iimits.

7.Txhua yam ua haujlwm tsis zoo

P1ease hu rau SAIRRAY tam sim ntawd,Yog tias pom muaj kev ua haujlwm tsis zoo ntawm X-ray raj sib dhos.

8. Pov tseg

X-ray raj sib dhos nrog rau lub raj muaj cov ntaub ntawv xws li roj thiab cov hlau hnyav uas ib puag ncig tus phooj ywg thiab pov tseg kom raug raws li kev cai lij choj hauv tebchaws yuav tsum tau lees paub.Kev pov tseg raws li kev lag luam hauv tsev lossis kev lag luam yog txwv tsis pub.Cov chaw tsim khoom muaj qhov kev paub txog kev xav tau thiab yuav siv lub X-ray raj sib dhos rov qab rau kev pov tseg.

Thov hu rau tus neeg muab kev pabcuam rau lub hom phiaj no.

Emission curves ntawm cathode

kev

Yog (A) Me Focal Spot

Cov Lus Qhia 1.2

Yog (A) Loj Focal Spot

Ib leeg thiab series loading

Cov xwm txheej: Tube Voltage Peb-theem

Stator Fais Fab Zaus 50Hz/60Hz

Ib (mA)

kev 4

t(s)

Ib (mA)

ywb 3

t(s)

Cua sov thiab txias nkhaus ntawm anode

  IEC60613

kev 5

Cua sov thiab txias nkhaus ntawm X-ray raj sib dhos

Tsev Thermal Yam ntxwv

kev 6

X-ray tube sib dhos dimensional kos duab

SWB 7110A

xoj 7

Lim los ua ke thiab hla ntu ntawm chaw nres nkoj

lus 27

Rotor Connector Wiring

lus 28

  • Yav dhau los:
  • Tom ntej:

  • Yam tsawg kawg nkaus Order Quantity: 1pc

    Nqe: Kev sib tham

    Ntim Cov Lus Qhia: 100pcs ib lub thawv lossis customized raws li qhov ntau

    Lub sij hawm xa tuaj: 1 ~ 2 lub lis piam raws li qhov ntau

    Cov nqe lus them nqi: 100% T / T ua ntej lossis WESTERN UNION

    Muaj peev xwm muab khoom: 1000pcs / hli

    Sau koj cov lus ntawm no thiab xa tuaj rau peb