Cov khoom no tau raug tsim thiab tsim nyob rau hauv kev pom zoo nrog cov cai hauv qab no, cov lus qhia thiab kev tsim qauv:
◆ Council Directive 93/42/EEC ntawm 14 Lub Rau Hli 1993 txog cov khoom siv kho mob(CE kos npe).
TS EN ISO 13485: 2016 Cov cuab yeej kho mob - Kev tswj hwm zoo - Cov kev xav tau rau kev tswj hwm
lub hom phiaj..
TS EN ISO 14971: 2012 Cov cuab yeej kho mob - Kev siv kev tswj hwm kev pheej hmoo rau cov cuab yeej kho mob (ISO 14971: 2007, Kho version 2007-10-01)
◆ TS EN ISO 15223-1: 2012 Cov cuab yeej kho mob - Cov cim siv nrog cov ntawv sau cov cuab yeej kho mob, daim ntawv lo thiab cov ntaub ntawv yuav tsum tau muab rau Ntu 1: Cov kev cai dav dav
◆International Electrotechnical Commission (IEC), cov qauv hauv qab no raug txiav txim siab tshwj xeeb.
Standard Reference | Cov npe |
EN 60601-2-54: 2009 | Cov khoom siv hluav taws xob kho mob - Ntu 2-54: Cov kev cai tshwj xeeb rau kev nyab xeeb thiab kev ua haujlwm tseem ceeb ntawm cov khoom siv xoo hluav taws xob rau hluav taws xob thiab xov tooj cua |
IEC 60526: Cov | High-voltage cable plug thiab socket txuas rau cov khoom siv kho mob X-ray |
IEC 60522: 1999 (TSI) | Kev txiav txim siab mus tas li pom ntawm X-ray raj sib dhos |
IEC 60613-2010 1 | Hluav taws xob, thermal thiab loading yam ntxwv ntawm rotating anode X-ray raj rau kev kuaj mob |
IEC 60601-1: 2006 | Cov khoom siv hluav taws xob kho mob - Ntu 1: Cov kev cai dav dav rau kev nyab xeeb thiab kev ua haujlwm tseem ceeb |
IEC 60601-1-3: 2008 | Cov cuab yeej siv kho mob hluav taws xob - Ntu 1-3: Cov kev cai dav dav rau kev nyab xeeb thiab kev ua haujlwm tseem ceeb - Cov Qauv Txheej Txheem: Kev tiv thaiv hluav taws xob hauv cov khoom siv kuaj X-ray |
IEC 60601-2-28: 2010 | Cov khoom siv hluav taws xob kho mob - Ntu 2-28: Cov kev cai tshwj xeeb rau kev nyab xeeb thiab kev ua haujlwm tseem ceeb ntawm X-ray raj sib dhos rau kev kuaj mob |
IEC 60336-2005 1 | Cov khoom siv hluav taws xob kho mob-X-ray raj sib dhos rau kev kuaj mob-Cov yam ntxwv ntawm cov chaw focal |
● Lub npe yog tsim raws li hauv qab no:
MWHX 7010 | Tube | A | High voltage socket nrog 90 degree kev taw qhia |
MWTX 70-1.0/2.0-125 | B | High voltage socket nrog 270 degree kev taw qhia |
Khoom | Specification | Txuj | |
Nominal input zog (s) ntawm lub anode | F 1 | F 2 | TS EN 60613 |
21kW (50/60Hz) | 42.5kW (50/60Hz) | ||
Anode tshav kub cia muaj peev xwm | 100 kJ (140 kHU) | TS EN 60613 | |
Qhov siab tshaj plaws txias peev xwm ntawm lub anode | 475 wb | ||
Thaum tshav kub kub cia muaj peev xwm | 9 00kj | ||
Max. nruam kub dissipation yam tsis muaj cua-cim | 180 W | ||
Cov khoom siv anodeAnode sab saum toj txheej khoom | Rhenium-Tungsten-TZM (RTM) Rhenium-Tungsten-(RT) | ||
Lub hom phiaj (Ref: reference axis) | 16 ° | IEC 60788 (ib.) | |
X-ray tube los ua ke inherent filtration | 1.5 hli Al / 75kV | IEC 60601-1-3: Kev qhia tshwj xeeb | |
Focal Spot nominal tus nqi | F1 (me me tsom) | F2 (loj tsom xam) | IEC 60336 (TSI) |
1.0 | 2.0 | ||
X-ray tube nominal voltageRadiographic Fluoroscopic | 125 kV ua 100kV ua | TS EN 60613 | |
Cov ntaub ntawv ntawm cathode cua sov Max. tam sim no Max voltage | ≈ / AC, <20 kHz | ||
F1 | F 2 | ||
5.1A ≈5.8~7.8v ua | 5.1 A ≈7.7~10.4 V | ||
Cov hluav taws xob tawg ntawm 150 kV / 3mA hauv 1m nrug | ≤1.0mGy/h | IEC 60601-1-3 | |
Qhov siab tshaj plaws hluav taws xob | 573 × 573mm ntawm SID 1m | ||
X-ray raj sib dhos hnyav | Kwv yees li. 18 kg |
txwv | Kev txwv kev ua haujlwm | Kev thauj mus los thiab kev txwv cia |
Ambient kub | Los ntawm 10℃rau 40℃ | Los ntawm - 20℃to 70℃ |
txheeb ze humidity | ≤ 75% | ≤ 93% |
Barometric siab | Los ntawm 70kPa mus rau 106kPa | Los ntawm 70kPa mus rau 106kPa |
1-theem stator
Test point | C-M | C-A |
Winding kuj | ≈18.0…22.0Ω | ≈45.0…55.0Ω |
Max.permissible khiav hauj lwm voltage (run-up) | 230V ± 10% Tawm tswv yim | |
Pom zoo ua haujlwm voltage (run-up) | 160V ± 10% Tawm tswv yim | |
Braking voltage | 70vdc ua | |
Khiav-on voltage nyob rau hauv raug | 80 vwm | |
Kev khiav hauj lwm voltage hauv fluoroscopy | 20V-40Vrms | |
Lub sijhawm ua haujlwm (nyob ntawm qhov pib qhov system) | 1.2s ib |
Ceeb toom rau interface nrog X-ray Generator
1. Vaj tse Repture
Tsis txhob tso tawm ntau tshaj lub zog rau X-ray raj sib dhos
Yog hais tias lub hwj chim input tshaj lub raj lub specifications, nws yuav ua rau lub anode overheat, lub raj iav tawg, thiab thaum kawg yuav ua rau muaj teeb meem loj vim overvoltage los ntawm vaporization ntawm cov roj nyob rau hauv lub vaj huam sib luag. Nyob rau hauv ib qho teeb meem tseem ceeb uas lub tsev ruptures vim overload, kev nyab xeeb thermal hloov yuav tsis muaj peev xwm tiv thaiv lub X-ray raj, txawm tias nws ua haujlwm.
*Lub tsev sealing qhov chaw rupture.
*Tib neeg raug mob nrog rau kub hnyiab vim yog roj kub khiav.
*Hluav taws kub sib tsoo vim yog lub hom phiaj flaming anode.
Lub tshuab hluav taws xob xoo hluav taws xob yuav tsum muaj kev tiv thaiv kev ua haujlwm uas tswj xyuas lub zog hluav taws xob kom nyob hauv cov raj tshwj xeeb.
2. Hluav taws xob poob siab
Txhawm rau kom tsis txhob muaj kev pheej hmoo ntawm hluav taws xob poob siab, cov cuab yeej no yuav tsum tsuas yog txuas nrog cov khoom siv tiv thaiv lub ntiaj teb.
3.Tsis muaj kev hloov kho ntawm cov cuab yeej no raug tso cai!!
Ceev faj rau kev sib txuas nrog X-ray Generator
1. Tshaj qhov ntsuas
Lub zog ntau dhau ntawm kev txhaj tshuaj ib zaug tuaj yeem ua rau X-ray tube sib dhos tsis ua haujlwm.Nws yog ib qho tseem ceeb kom ua tib zoo tshuaj xyuas cov ntaub ntawv cov ntaub ntawv thiab ua raws li cov lus qhia tshwj xeeb kom tsis txhob muaj kev puas tsuaj.
2.Kev lim dej tas mus li
Cov kev cai lij choj hais txog tag nrho cov kev pom uas yuav tsum tau ua thiab qhov tsawg kawg nkaus nyob deb ntawm X-ray tsom xam thiab tib neeg lub cev.
Thav yuav tsum ua raws li txoj cai.
3.Kev Nyab Xeeb Thermal Hloov
X-ray tube sib dhos muaj kev nyab xeeb thermal hloov los txwv tsis pub muaj zog ntxiv thaum lub raj vaj tse mus txog qhov kub thiab txias(80℃)ntawm switch-open.
Qhov hloov tsis pom zoo txuas stator coil hauv series Circuit Court.
Txawm hais tias qhov hloov pauv ua haujlwm, tsis txhob tig lub zog hluav taws xob tawm. Lub tshuab cua txias yuav tsum tau qhib yog tias siv nrog lub kaw lus.
4.Kev ua haujlwm tsis zoo
X-ray raj sib dhos tuaj yeem ua haujlwm tsis zoo lossis ua tsis tiav, tsim kom muaj teeb meem loj. Nws yog ib qho tseem ceeb uas yuav tsum muaj txoj kev npaj xwm txheej ceev los tiv thaiv thiab daws txhua yam teeb meem uas yuav tshwm sim los ntawm qhov kev pheej hmoo no.
5. Daim ntawv thov tshiab
Yog tias koj npaj yuav siv cov khoom no hauv daim ntawv thov tshiab uas tsis tau teev tseg hauv daim ntawv no, lossis yog tias koj npaj yuav siv lwm hom X-ray tshuab hluav taws xob, thov hu rau peb kom paub meej tias muaj kev sib raug zoo thiab muaj.
1 .X-ray Radiationkev tiv thaiv
Cov khoom no ua tiav cov kev cai ntawm IEC 60601-1-3.
X-ray raj sib dhos no tso tawm cov hluav taws xob xoo hluav taws xob hauv kev ua haujlwm.Tsuas yog cov neeg ua haujlwm tsim nyog thiab cov neeg ua haujlwm tau txais kev tso cai ua haujlwm X-ray raj sib dhos.
Cov teebmeem physiologic cuam tshuam yuav ua rau muaj kev phom sij rau tus neeg mob, kev tsim khoom yuav tsum tau txais kev tiv thaiv kom tsis txhob muaj hluav taws xob ionization.
2. Dielectric 0 il
X-ray raj sib dhos muaj dielectric 0il muaj rau kev ruaj ntseg siab. Raws li nws yog tshuaj lom rau tib neeg noj qab haus huv,yog tias nws raug rau thaj tsam tsis txwv,nws yuav tsum muab pov tseg raws li kev cai hauv zos.
3 .Operation Atmosphere
X-ray raj sib dhos tsis raug tso cai siv nyob rau hauv cov huab cua ntawm cov nplaim taws lossis cov roj corrosive ·
4.Kho lub raj tam sim no
Nyob ntawm cov xwm txheej ua haujlwm,cov yam ntxwv filament yuav raug hloov.
Qhov kev hloov pauv no tuaj yeem ua rau muaj kev cuam tshuam ntau dua rau X-ray raj sib dhos.
Txhawm rau tiv thaiv X-ray tube los ua kom puas,kho lub raj tam sim no tsis tu ncua.
Dhau li ntawm thaum X-ray raj muaj teeb meem arcing hauv iblsiv sijhawm ntev,yuav tsum tau kho lub raj tam sim no.
5.X-ray Tube Vaj Tse Kub
Tsis txhob kov ntawm X-ray raj lub tsev nyob saum npoo tsuas yog tom qab ua haujlwm vim qhov kub thiab txias.
Nyob twj ywm X-ray raj kom txias.
6.Txoj kev txwv
Ua ntej siv,thov paub meej tias qhov xwm txheej ib puag ncig yog nyob rau hauv kev khiav hauj lwm Iimits.
7.Txhua yam ua haujlwm tsis zoo
P1ease hu rau SAIRRAY tam sim ntawd,Yog tias pom muaj kev ua haujlwm tsis zoo ntawm X-ray raj sib dhos.
8. Pov tseg
X-ray raj sib dhos nrog rau lub raj muaj cov ntaub ntawv xws li roj thiab cov hlau hnyav uas ib puag ncig tus phooj ywg thiab pov tseg kom raug raws li kev cai lij choj hauv tebchaws yuav tsum tau lees paub.Kev pov tseg raws li kev lag luam hauv tsev lossis kev lag luam yog txwv tsis pub.Cov chaw tsim khoom muaj qhov kev paub txog kev xav tau thiab yuav siv lub X-ray raj sib dhos rov qab rau kev pov tseg.
Thov hu rau tus neeg muab kev pabcuam rau lub hom phiaj no.
Yog (A) Me Me Focal Spot
Yog (A) Loj Focal Spot
Cov xwm txheej: Tube Voltage Peb-theem
Stator Fais Fab Zaus 50Hz/60Hz
IEC60613
Tsev Thermal Yam ntxwv
SRMWHX 7010A
SRMWHX 7010B
Lim los ua ke thiab hla ntu ntawm chaw nres nkoj
Rotor Connector Wiring
Yam tsawg kawg nkaus Order Quantity: 1pc
Nqe: Kev sib tham
Ntim Cov Lus Qhia: 100pcs ib lub thawv lossis customized raws li qhov ntau
Lub sij hawm xa tuaj: 1 ~ 2 lub lis piam raws li qhov ntau
Cov nqe lus them nqi: 100% T / T ua ntej lossis WESTERN UNION
Muaj peev xwm muab khoom: 1000pcs / hli